Endoluminal Therapy for Treatment of Gastroesophageal Reflux Disease

A336131_1_En_7_Fig1b_HTML.jpg



Fig. 7.1
The EsophyX® device: first- and second-generation devices (courtesy of EndoGastric Solutions, Inc. Redmond, WA, USA). (a1, a2) The device currently used (©2014 EndoGastric Solutions, Inc.). (b1, b2) The new generation device (©2014 EndoGastric Solutions, Inc.)



The procedure is performed by two operators: one controls the device and the other one operates the endoscope.

The device is inserted transorally with the patient in the left lateral or supine position, under general anesthesia. Hypopharyngeal perforation has been reported in this phase of the procedure if the device is introduced without an adequate caution; in difficult cases, the device can be gently rotated to pass the upper esophageal sphincter.

Once into the stomach, air or CO2 is insufflated to distend the gastric cavity and permit an adequate vision of the gastric fundus and EGJ; CO2 is preferable, because it leads to a faster and more sustained gastric insufflation and induces less discomfort to patients. With the endoscope placed in retroflexion position, the lesser curve is located at the 12 o’clock position and the greater curve at the 6 o’clock in the patient placed in left decubitus. Once the tissue mold is retroflexed, it is closed against the EsophyX® device, rotated to 11 or 1 o’clock position (lesser curve), and pulled back to place its tip just inside the esophageal lumen. At this point, the helical screw is advanced to engage the tissue under direct vision just below the Z line, the shaft of the device is advanced caudally, the tissue mold is opened, and the helical screw cable freed from the tissue mold. Then, a tension is applied to helical retractor, while a slight opening and closing of the tissue mold allow the fundus to slide through the tissue mold; in this phase the stomach is being desufflated. Failure to desufflate the stomach during this phase of the procedure limits the size of the fundoplication.

After completing this maneuver, both helical retractor and tissue mold are locked in place; suction is applied to the tissue invaginator for approximately half a minute, and the device is then advanced caudally into the stomach, which has been re-insufflated. The latter maneuver ensures that esophagogastric plication is performed in an intra-abdominal position and reduces hiatal hernia, when present.

Plication is carried out by deploying multiple polypropylenes, H-shaped fasteners advanced over two stylets, one anterior and the other posterior. The fastener deployment process initiates on the far posterior and anterior sides of the esophagogastric valve adjacent to the lesser curvature and then is extended to the greater curvature by rotating the tissue mold axially to slide the stomach over the esophagus, resulting in circumferential tightening and a new valve circumference of >240°. The stylet is advanced under direct endoscopical vision through the tissue molded until its tip is seen by the operator. The fastener is then advanced over the stylet and deployed to create a serosa-to-serosa plication. Once the tip of the fastener becomes visible at the tissue mold, the stylet is pulled back while the fastener is maintained in place; by this way, the leading leg of the fastener is derailed and the fastener is deployed. Fourteen fasteners allowing seven plications are needed to construct a satisfactory circumferential gastroesophageal valve; however, the higher is the number of fasteners deployed, the more continent is the newly created valve. Details of the EsophyX® device’s technique are shown in Fig. 7.2.

A336131_1_En_7_Fig2_HTML.jpg


Fig. 7.2
Schematic representation of the procedure with EsophyX® device (Courtesy of EndoGastric Solutions Inc. Redmond, WA, USA). (a) The EsophyX® device enters the esophagus through the mouth and is positioned at the gastroesophageal junction; (b) the device wraps the fundus around the distal esophagus and fastens a tissue fold; this step is then repeated multiple times to reconstruct a robust, tight valve (c) (©2014 EndoGastric Solutions, Inc.)

Endoscopic pre- and post-procedural findings are reported in Fig. 7.3.

A336131_1_En_7_Fig3_HTML.gif


Fig. 7.3
Endoscopic views of the gastroesophageal valve before and immediately after the TIF procedure by EsophyX® device (authors’ case). (a) The gastroesophageal valve: before the procedure with the EsophyX® device; (b) the “Bell Roll” maneuver to create the new gastroesophageal valve; (c) the gastroesophageal valve: immediately after the procedure with the EsophyX® device; (d) the gastroesophageal valve: 6 months after the procedure

Besides the standard procedure, two modified techniques have been reported over time to create the fundoplication. The technique we used in the last years engages the tissue by advancing the helical screw just below the Z line on the far posterior and anterior sides of the esophagogastric valve adjacent to the lesser curvature (11 and 1 o’clock position). Before inserting the stylet, a torque is applied by rotation (clockwise and counterclockwise at 11 and 1 o’clock, respectively) of the tissue mold locked; such a maneuver allows part of the fundus to rotate around the esophageal wall and more tissue to be engaged by the stylet. Four fasteners for each site are deployed at 1 and 11 o’clock and two fasteners for each site in the middle part of the valve, at 4, 6, and 8 o’clock position, to reinforce and prolong caudally the plication. This technique increased by 30 % the success rate of the procedure. With the standard TIF technique, 11/27 patients (40.7 %) didn’t take PPI therapy at 12 months; with the application of the rotational TIF technique, 14/22 patients (63.6 %) were full responders.

Bell R. et al. have developed a rotational fundoplication, the so-called Bell Roll maneuver [38]. The helical retractor is engaged at 12 o’clock, and the tissue mold is placed at 6 o’clock; then the tissue mold locked is rotated toward the lesser curve by a radial motion of the handle of the device to the 12 o’clock position. This maneuver rolls the fundus over and around the distal esophagus to the 1 o’clock position.

At the end of the plication, an immediate endoscopy is performed to evaluate the pharynx, the esophageal lumen, the gastric fundus, and the fundoplication.



7.3.3 Transoral Fundoplication by MUSETM System


The MUSE™ system includes the endostapler and a console connected with the endostapler, containing a controller for the camera, ultrasonic range finder and various sensors, a pump for insufflation and irrigation, a suction system, and power and controls for the LED.

The endostapler has:


  1. (a)


    A handle, wherein controls are located

     

  2. (b)


    An insertion tube 15.5 mm in diameter, 66 cm long, containing the suction, insufflation/irrigation channels, and electrical and mechanical cables which operate the device

     

  3. (c)


    A rigid section 66 mm in length that contains the cartridge. Each cartridge contains five standard 4.8-mm titanium staples, the ultrasound mirror, one alignment pin funnel, and two anvil screw funnels

     

  4. (d)


    The distal tip, similar to that of an endoscope, with suction, irrigation, illumination (via LED), and visualization (via miniature camera) capabilities

     
The anvil, alignment pin, anvil screw, and ultrasound are all designed to ensure proper alignment and positioning of the device during stapling. The distal tip may be articulated in one direction to align with the rigid section and cartridge, with a bending radius of 26 and 40 mm.

Details of the device are illustrated in Fig. 7.4.

A336131_1_En_7_Fig4_HTML.gif


Fig. 7.4
The Medigus Surgical Ultrasonic Endostapler system, MUSE™ (Courtesy of Medigus Ltd., Omer, Israel). (a) The MUSE™ system (© All rights reserved to Medigus Ltd. 20082015); (b) the console connected with the endostapler, containing a controller for the camera, ultrasonic range finder, and various sensors (bending angle, bending force, alignment pin, anvil screws, gap) (© All rights reserved to Medigus Ltd. 20082015)

The procedure can be performed by one operator in experienced hands. The patient is placed in the supine position, under general anesthesia with endotracheal intubation. Positive end-expiratory pressure (PEEP) of at least 5 mmHg (7.5-cm H2O) is administered. After a preliminary endoscopic assessment of the esophagus and stomach and once no contraindications are found, an overtube is placed. Then, the endostapler is inserted transorally through the overtube and gently advanced into the stomach under direct vision; passing the rigid section across the pharyngoesophageal junction may encounter some resistance. In order to avoid applying excessive force and risk to injure the esophagus, the overtube may be withdrawn approximately 5 cm and then advanced with the endostapler as a unit. This maneuver can be repeated until the system reaches the esophageal midbody. Flexing the neck may make passage easier.

Once into the stomach, distended by insufflation of air or CO2, the stapler is advanced until the tip is approximately 5 cm past the EGJ and then retroflex by 180° to obtain an adequate vision of the gastric fundus and EGJ to select stapling location.

The most important stapling location is the leftmost location, which is typically performed first. This is the anchoring point for the fundus and should be placed as far to the left of the esophagus as possible. At times, depending on anatomy, it may be easier to perform the first stapling in a more central location. The additional stapling locations should be within 60–180° as long as the rightmost stapling should not be done on the lesser curve, because stapling in the lesser curve may attach the antrum to the esophagus and open the esophagogastric junction rather than close it. Additional staplings may be placed between the leftmost and rightmost. Once the correct location for stapling has been identified, all the procedures are performed under ultrasound guidance. Subsequent phases of the procedure include clamping tissue, deploying alignment pin, advancing anvil screw, stapling, and retrieving anvil screws [29]. Details of the MUSETM device’s technique are shown in Fig. 7.5. Endoscopic pre- and post-procedural findings after TIF with MUSETM system are reported in Fig. 7.6.

A336131_1_En_7_Fig5_HTML.gif


Fig. 7.5
Schematic representation of the Medigus Ultrasonic Surgical Endostapler (MUSE™) procedure (Courtesy of Medigus Ltd., Omer, Israel). (a) The endostapler is inserted transorally through the overtube and gently advanced into the stomach under direct vision; (b) once in the stomach, distended by insufflation of air or CO2, the stapler is advanced until the tip is approximately 5 cm past the EGJ and then retroflexed 180° to give adequate vision of the gastric fundus and EGJ to select the stapling location. The tissue is clamped and stapled under ultrasonic guidance; (c) this step is then repeated at least twice to reconstruct a robust, tight valve. Additional stapling locations should be within 60–180° of the valve circumference. EGJ, esophagogastric junction (© All rights reserved to Medigus Ltd. 20082015)


A336131_1_En_7_Fig6_HTML.gif


Fig. 7.6
Endoscopic views of the gastroesophageal valve before and after the TIF procedure with the Medigus Ultrasonic Surgical Endostapler (MUSE™) (authors’ case). (a) The gastroesophageal valve: before the TIF procedure with the MUSE™ system; (b) the gastroesophageal valve: immediately after the TIF procedure by MUSE™ system; (c) the gastroesophageal valve: 6 months after the TIF procedure by MUSE™ system


7.3.4 Postoperative Care


Antiemetic prophylaxis with at least two drugs (according to the ASA recommendations for interventions with high risk of post-procedure nausea and vomiting) and full muscle relaxation throughout the procedure are mandatory for TIF. Antiemetic prophylaxis is maintained i.v. for 24 h, while broad-spectrum antibiotic therapy is maintained i.v. for 48 h and then by oral route over a 5-day period.

Almost all patients complain of transient pharyngeal irritation, as a result of insertion and manipulation of the device, and some have mild to moderate epigastric pain in 6 h after the procedure. Pain persisting for 2–4 days may require analgesics and should be considered for esophageal or gastric leak; CT scan and hydrosoluble contrast X-ray investigation should be carried out in these cases. Dysphagia and gas bloating are generally not reported by patients. White blood cell count may be slightly increased after the procedure. At discharge, patients are instructed to follow a liquid diet for the first 2 weeks and a soft diet for the next 4 weeks. PPIs are discontinued 7 days after the procedure. Patients are also asked to refrain from vigorous exercise for 4 weeks.



7.4 Outcomes


To date, 20 prospective studies and one retrospective study, most of them observational and carried out in a limited number of patients, have been published on short- and medium-term follow-up (1–3 years) after TIF using EsophyX® device [3858]. One study evaluated patients’ outcomes up to 6 years after the procedure [56]. Sixteen studies assessed symptoms by means of the GERD health-related quality of life (HRQL); 11 evaluated pre- and post-procedure pH ± impedance recordings. A multicenter prospective study compared the efficacy of TIF versus omeprazole in a randomized controlled trial [59].

Overall, in 17 studies TIF was proven to discontinue antireflux medications or markedly decrease their dose; three studies arose concerns about the effectiveness of the procedure [45, 46, 49].

In the observational, nonrandomized studies, 6- and 12-month outcomes after TIF showed that 75–84 % and 53–85 % of patients had either discontinued PPI use or halved the dose of PPI therapy. Normalization of esophageal acid exposure, in terms of total acidic refluxes, number of refluxes, and DeMeester score, was reported in 37–89 % of patients.

Twenty-four months after TIF, daily high-dosage PPI dependence was eliminated in 75–93 % of patients. Endoscopic findings comparing fundoplication immediately after the procedure and 2 years later are reported in Fig. 7.7. In the two series reporting a 3-year follow-up, persistent discontinuation of daily PPI ranged from 74 to 84 % of cases [54, 56].

A336131_1_En_7_Fig7_HTML.gif


Fig. 7.7
Endoscopic views of the gastroesophageal valve immediately after and 24 months after the TIF procedure with EsophyX® device (authors’ case). (a) The gastroesophageal valve: immediately after the TIF procedure with EsophyX® device; (b) the gastroesophageal valve: 24 months after the TIF procedure with EsophyX® device

Only one study evaluated outcomes 6 years after TIF in 14/50 patients who underwent the procedure. High-dosage PPI dependence was eliminated in 86 % of patients, and approximately half of them completely stopped PPI use [56], providing evidence of the long-lasting effect of TIF on symptoms and PPI usage. Results are summarized in Fig. 7.8. Unsuccessful outcomes of TIF occurred mainly between 6 and 12 months after the procedure, while between 12 and 36 months, the results did not substantially differ. Six-year results were substantially similar to those reported at 36 months. These findings show that an appropriate patient selection plays a pivotal role in achieving clinical success after TIF and confirm that factors negatively affecting postoperative outcomes play a role early in the postoperative period in most patients. Operator’s experience plays an important role in TIF outcomes, too. All TIF failures observed in our series occurred in patients who underwent the procedure early in the operator’s learning curve. A retrospective study in 124 unselected patients carried out in two community hospitals and reporting, respectively, 75 and 80 % of patients free of typical and atypical GERD symptoms over a mean follow-up of 7 months confirmed that the operator’s experience plays a major role in successful outcomes [52].

A336131_1_En_7_Fig8_HTML.gif


Fig. 7.8
Symptomatic responses 6 months and 1–6 years after TIF with EsophyX® device, classified according to proton pump inhibitor (PPI) use. Patients were grouped as complete responders (who completely stopped using PPI) or partial responders (who halved the previous PPI dose) and nonresponders (who still used the pre-TIF PPI dose): 12 months versus 6 months after TIF P = 0.8; 24 versus 12 months P = 0.4; 36 versus 24 months P = 0.7; 4 years versus 36 months P = 1.0; 5 versus 4 years P = 1.0; 6 versus 5 years P = 1.0

Only three prospective, randomized controlled trials have been so far published. Two assessed the 6-month efficacy of TIF versus omeprazole therapy: one showed TIF more effective than PPI therapy in treating regurgitation and extraesophageal symptoms (97 % vs 50 % of patients, respectively; P = 0.006) [58]; the second one proved at intention-to-treat analysis TIF more effective than PPI in eliminating GERD symptoms (67 % vs 45 %, respectively; P = 0.023) [59]. These data show different outcomes and require additional randomized studies to clarify the efficacy of TIF in treating GERD. The third study compared 3- and 12-month results of TIF and Nissen fundoplication, showing TIF as effective and safe as Nissen fundoplication but with significantly lower hospital stay (2.9 ± 0.8 days vs 6.4 ± 0.7 days, respectively; P < 0.0001) [60].

Outcomes of TIF, with regard to the effects on PPI usage, are reported in Tables 7.1 and 7.2.


Table 7.1
Symptomatic responses after transoral incisionless fundoplication with the EsophyX® device





















































































































Follow-up

6 months

12 months

24 months

36 months

6 years

Cadière et al. [39]

2008


85 %




Cadière et al. [41]

2009



93 %



Testoni et al. [42]

2010

82 %

76 %




Velanovich et al. [44]

2010

79 %





Repici et al. [45]

2010

55 %

47 %




Demyttenaere et al. [43]

2010


53 %




Hoppo et al. [46]

2010


42 %




Barnes et al. [52]

2011

93 %





Bell et al. [47]

2011

75 %





Ihde et al. [48]

2011

76 %





Trad et al. [51]

2012


82 %




Testoni et al. [53]

2012



75 %

75 %


Petersen et al. [50]

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Aug 23, 2017 | Posted by in ABDOMINAL MEDICINE | Comments Off on Endoluminal Therapy for Treatment of Gastroesophageal Reflux Disease

Full access? Get Clinical Tree

Get Clinical Tree app for offline access